E-LITE Electro Surgical Generator
Compact & Lightweight Electrosurgical Generator for Open and Laparoscopic Surgeries
The E-LITE Electro Surgical Generator from the LITE Series is designed for small and general surgical procedures where compact size, mobility and dependable electrosurgical performance are important.
Its lightweight construction makes the unit easy to mobilize and hand-carry. The generator uses an advanced multiple-microcontroller-based design and is intended for both open and laparoscopic surgeries.
With monopolar and bipolar operating modes, tissue monitoring, patient plate contact monitoring, self-test functionality and intuitive controls, E-LITE provides essential electrosurgical capabilities in a compact platform.
Efficient Electrosurgical Performance in a Compact Design
E-LITE features a feather-touch membrane keyboard for convenient selection of power, mode, program and footswitch functions.
The system includes two monopolar sockets and one bipolar socket, supporting hand-switch and foot-pedal operation without frequent changeover. Its user-friendly interface is designed to simplify program and mode selection during surgical procedures.
The generator is also compatible with universal accessories used for monopolar and bipolar surgical applications.
Key Features of E-LITE
- Compact and lightweight construction for easy mobility and hand-carrying
- Designed for small and general surgical procedures
- Multiple microcontroller-based electrosurgical generator
- Suitable for open and laparoscopic surgeries
- Feather-touch membrane keyboard
- Convenient power, mode, program and footswitch selection
- Two monopolar and one bipolar sockets
- Hand-switch and foot-pedal operation
- Compatibility with universal monopolar and bipolar accessories
- Automatic self-test during startup
- Restoration of previously activated power settings following power interruption
- Compatibility with ALAN wired and wireless footswitches
- Multiple cutting modes with regulated wattages
- Isolated monopolar and bipolar outputs
- Tissue impedance monitoring with dynamic HF output adjustment
- HF power adjustment from the sterile field using the hand-switch pencil
- Automatic power stoppage with audio-visual indication during a detected problem
- HF current leakage monitoring and control below 10 µA, as stated in the brochure
- PPCM patient contact-quality monitoring
- Real-time graphical representation of patient contact quality
- Support for ALAN's capacitive patient plate
- Convection cooling system
- Designed to operate in lower-voltage situations without requiring a voltage stabilizer
Monopolar Cutting Modes
E-LITE provides three monopolar cutting modes for different surgical requirements.
| Mode | Maximum Output | Load | CF |
|---|---|---|---|
| Pure | 300 W | 300 Ohms | 1.5 |
| Lapcut | 200 W | 300 Ohms | 1.5 |
| Blend | 100 W | 300 Ohms | 2.5 |
Monopolar Coagulation Modes
| Mode | Maximum Output | Load | CF |
|---|---|---|---|
| Fulgurate | 100 W | 500 Ohms | 7 |
| Desiccate | 100 W | 500 Ohms | 5.5 |
Bipolar Coagulation Modes
| Mode | Maximum Output | Load | CF |
|---|---|---|---|
| Cut | 80 W | 100 Ohms | 1.5 |
| Macro | 80 W | 100 Ohms | 1.5 |
Electrical Specifications
| Specification | Details |
|---|---|
| Input Supply | 230 VAC ±20%, 50 Hz |
| Power Consumption | 750 VA Max |
| Input Fuse | 3.15A T |
| Maximum Power Output | 300 Watts |
Safety Standards
According to the manufacturer brochure, the E-LITE basic construction is designed according to:
- IEC 60601-1
- IEC 60601-1-2
- IEC 60601-2-2
Classification & Technical Details
| Parameter | Specification |
|---|---|
| Protection Class | I |
| MDD Class | II b |
| Pollution Degree Class | II |
| Overvoltage Category | II |
| Unit Type | CF |
| Outputs | Floating Output |
| Electrical Potential Balancing | Indicated by Symbol |
| PER | 98 |
| Dimensions | 285 mm (W) × 150 mm (H) × 310 mm (D) |
| Weight | Maximum 4.2 kg |
Compact Size for Easy Mobility
With dimensions of only 285 × 150 × 310 mm and a maximum weight of 4.2 kg, the E-LITE is designed to provide convenient mobility within appropriate surgical environments.
Its compact form makes it particularly suitable where a smaller electrosurgical generator is preferred without sacrificing essential monopolar and bipolar capabilities.
Smart Tissue Monitoring Technology
E-LITE incorporates tissue monitoring technology that measures tissue impedance and dynamically varies the HF output.
This helps the generator respond to changing tissue conditions during electrosurgical procedures. The HF output can also be adjusted using the hand-switch pencil from the sterile field.
Patient Plate Contact Monitoring – PPCM
The E-LITE generator includes PPCM – Patient Plate Contact Monitoring, which monitors patient contact quality and provides a real-time graphical representation.
The system also supports ALAN's special capacitive patient plate.
Automatic Safety Monitoring
The system is designed to automatically stop power output and provide an audio-visual indication when a problem is detected.
The brochure also specifies an HF current leakage monitoring and controlling system that supervises leakage and maintains it below 10 µA.
Self-Test & Power Setting Memory
At every startup, the generator performs a self-test to check system health and operational integrity.
In the event of a power interruption, E-LITE restores the previously activated power settings, helping maintain operational convenience.
Wired & Wireless Footswitch Compatibility
E-LITE is compatible with ALAN wired and wireless footswitches, offering additional flexibility and convenience during surgical procedures.
Operating Conditions
The manufacturer specifies the following operating conditions:
| Parameter | Range |
|---|---|
| Ambient Temperature | +10°C to +40°C |
| Relative Humidity | 30% to 75%, non-condensing |
| Atmospheric Pressure | 700 to 1060 millibars |
Transport & Storage Conditions
| Parameter | Range |
|---|---|
| Ambient Temperature | -30°C to +65°C |
| Relative Humidity | 0% to 90%, non-condensing |
| Atmospheric Pressure | 500 to 1060 millibars |
Designed for Practical Surgical Applications
The E-LITE Electro Surgical Generator combines compact size with a maximum output of 300 Watts, multiple monopolar and bipolar modes, tissue monitoring and patient contact-quality monitoring.
Its combination of portability, intuitive controls and essential electrosurgical functionality makes it suitable for the small and general surgical applications described in the manufacturer brochure.
Enquire About E-LITE Electro Surgical Generator
Looking for a Compact Electrosurgical Generator?
Connect with Cygnus Tech Lab to learn more about the E-LITE Electro Surgical Generator, technical specifications and product requirements.